Big changes
This commit is contained in:
@@ -0,0 +1,95 @@
|
||||
; ===============================================================================================================
|
||||
; This name must match the name specified in the Instrument.xml that is associated with the power supply system
|
||||
[STE_POWER_SUPPLY_SYSTEM]
|
||||
SCPI_DEF_FILEPATH = .\InstrumentConfig\KEYSIGHT_SCPI_DEF.ini
|
||||
ETHERNET_ADDRESS = localhost
|
||||
ETHERNET_PORT = 5025
|
||||
MODULE_DEFINITION = UUT_REF_3_3V, STE_PVM_5V, STE_GU_INTERFACE_RELAYS_25V, STE_GU_INTERFACE_RF_INTERFACE_5V
|
||||
; 0 means no coupled modules.
|
||||
; couple means turning on/off any one of the module, turns on/off the others
|
||||
COUPLED_MODULES = STE_PVM_5V, STE_GU_INTERFACE_RELAYS_25V, STE_GU_INTERFACE_RF_INTERFACE_5V
|
||||
; 0 means no grouped modules.
|
||||
; group means turning combining 2 or more modules thus acting as one module
|
||||
GROUPED_MODULES = 0
|
||||
INTERFACE = ETHERNET
|
||||
|
||||
[STE_POWER_SUPPLY_SYSTEM.UUT_REF_3_3V]
|
||||
INDEX = 1
|
||||
OCP = 2.0
|
||||
OVP = 4.0
|
||||
VOLTAGE_SETPOINT = 3.3
|
||||
|
||||
MIN_VOLTAGE = 3.0
|
||||
MAX_VOLTAGE = 3.75
|
||||
MAX_CURRENT = 2.0
|
||||
MIN_CURRENT = -0.25
|
||||
|
||||
[STE_POWER_SUPPLY_SYSTEM.STE_PVM_5V]
|
||||
INDEX = 2
|
||||
OCP = 2.0
|
||||
OVP = 8.0
|
||||
VOLTAGE_SETPOINT = 5.0
|
||||
|
||||
MIN_VOLTAGE = 4.0
|
||||
MAX_VOLTAGE = 5.5
|
||||
MAX_CURRENT = 1.5
|
||||
MIN_CURRENT = -.25
|
||||
|
||||
[STE_POWER_SUPPLY_SYSTEM.STE_GU_INTERFACE_RELAYS_25V]
|
||||
INDEX = 3
|
||||
OCP = 3.0
|
||||
OVP = 34.0
|
||||
VOLTAGE_SETPOINT = 25.0
|
||||
|
||||
MIN_VOLTAGE = 22.0
|
||||
MAX_VOLTAGE = 30.0
|
||||
MAX_CURRENT = 3.0
|
||||
MIN_CURRENT = -0.25
|
||||
|
||||
[STE_POWER_SUPPLY_SYSTEM.STE_GU_INTERFACE_RF_INTERFACE_5V]
|
||||
INDEX = 4
|
||||
OCP = 5.0
|
||||
OVP = 6.5
|
||||
VOLTAGE_SETPOINT = 5.0
|
||||
|
||||
MIN_VOLTAGE = 4.0
|
||||
MAX_VOLTAGE = 6.3
|
||||
MAX_CURRENT = 4.5
|
||||
MIN_CURRENT = -.25
|
||||
|
||||
; ===============================================================================================================
|
||||
; This name must match the name specified in the Instrument.xml that is associated with the power supply system
|
||||
[UUT_POWER_SUPPLY_SYSTEM]
|
||||
SCPI_DEF_FILEPATH = .\InstrumentConfig\KEYSIGHT_SCPI_DEF.ini
|
||||
ETHERNET_ADDRESS = localhost
|
||||
ETHERNET_PORT = 5026
|
||||
MODULE_DEFINITION = UUT_P20V, UUT_N20V
|
||||
; 0 means no coupled modules.
|
||||
; couple means turning on/off any one of the module, turns on/off the others
|
||||
COUPLED_MODULES = 0
|
||||
; 0 means no grouped modules.
|
||||
; group means turning combining 2 or more modules thus acting as one module
|
||||
GROUPED_MODULES = 0
|
||||
INTERFACE = ETHERNET
|
||||
|
||||
[UUT_POWER_SUPPLY_SYSTEM.UUT_P20V]
|
||||
INDEX = 1
|
||||
OCP = 1.5
|
||||
OVP = 22.0
|
||||
VOLTAGE_SETPOINT = 20.0
|
||||
|
||||
MIN_VOLTAGE = 19.0
|
||||
MAX_VOLTAGE = 22.75
|
||||
MAX_CURRENT = 2.3
|
||||
MIN_CURRENT = -0.25
|
||||
|
||||
[UUT_POWER_SUPPLY_SYSTEM.UUT_N20V]
|
||||
INDEX = 2
|
||||
OCP = 1.8
|
||||
OVP = 22.5
|
||||
VOLTAGE_SETPOINT = 20.0
|
||||
|
||||
MIN_VOLTAGE = 18.0
|
||||
MAX_VOLTAGE = 23.0
|
||||
MAX_CURRENT = 2.5
|
||||
MIN_CURRENT = -.25
|
||||
Reference in New Issue
Block a user